RECIRCULATION TOP REACTOR SAMPLER - Sampling Systems
Liquid / Slurries > Reactor Crystalliser Hydrogenator > Recirculation Top Reactor Sampler

Extracts a liquid sample from reactors and vessels via a dip pipe and re-circulation loop.
- Integrated pH / conductivity measurement.
- Seed & slurry charging option.
- Ideal for PAT applications – In-Situ & Extractive sampling.
- Sample is representative.
- Robust system for lower maintenance costs.
Operation & Cleaning
A tubular diaphragm pump is used to draw the sample up the dip pipe, around the re-circulation loop and return it to the vessel.
- The liquid to be sampled is re-circulated for a few minutes to ensure a representative sample; pH measurement can take place at this point.
- The sampling valve is either manually or pneumatically actuated to allow sample flow into a sampling bottle.
- Straight-through pumps design allows a full nitrogen blow-through to clean and dry the system after every sample
Pump design
- Handles viscous and abrasive products
- Can run dry without damage
- 10,000 hours maintenance free
- Capable of passing up to 6 mm diameter solids (up to 50% crystal concentration)
- Self priming, up to 3m height
Technical Features
- Unique compact FlowCell allows sampling, in-line measurement and large sightglass with no dead space
- Air operated tubular diaphragm pump.
- Fully Teflon Lined re-circulation loop, sampling valve and solvent flush valve.
- Borosilicate sample bottle - ISO GL 45 standard bottle connection (special threads available on request)
Ratings
- Pressure: 0 to 10 bar.
- Temperature: -15° to 180°C.
- All process contact materials FDA approved
- Atex II 2c T180
- CE marked
Options
- Inline pH probe and transmitter.
- Contained seed charging device
- Needle Septum Block for contained transfer of sample to sample bottle.
- Sample bottle connection can be machined to suit customer supplied bottle.
- Glovebox for high containment applications.
- Can be mounted on a mobile skid.



|